| Item number: 163IE-6272846 Manufacturer no.: 500 801 EAN/GTIN: n/a |
| |
|
| | |
| Flame Photometer FF-200 Single channel, low temperature flame photometer for use in industry, medicine and education. Can be operated with propane, butane, natural gas or liquid gas (LPG). With built-in electronic flame failure safety system for operational safety.
-Model FF-200D-I incl. Na, K, Ca, Ba and Li filters -Model FF-200D-R incl. Na, K and Li filters -3 year warranty
Supplied with: Na, K, Li filters (model FF-200D-R), additionally Ba and Ca filter (model FF-200D-I), connecting hoses and clips, compressor plug, drain trap, fuses and operating/service instructions Specifications Range FF-200D-I: Range FF-200D-R: Limits of detection Na/K: Li: Ca: Ba: 0 ... 199.9 ppm 120 ... 160 ppm Na, 0 ... 10 ppm K
< 0.2 ppm < 0.25 ppm < 15 ppm < 30 ppm Reproducibility: < 1 % (Coefficient of variation for 20 consecutive samples using 10 ppm Na set to a reading of 50) Linearity: < 2 % (when concentration of 3 ppm Na/K and 5 ppm Li are set to a reading of 100) Specificity: Interference from Na, K and Li equal in concentration to test element will be < 0.5 % Recorder output: Nominal 1.00 V for a reading of 100.0 Air: 6 l/min at 1 Kg/cm2 /14psi (oil and moisture free pressurised air) Dimensions (W x D x H): Weight: 420 x 360 x 300 mm 8 kg Power supply: 190 ... 250 V, 50/60 Hz
Type FF-200D-R |
| | |
| | | |
| | | |
| |